ISO 3924:2010 pdf free download – Petroleum products一Determination of boiling range distribution一Gas chromatography method

02-10-2022 comment

ISO 3924:2010 pdf free download – Petroleum products一Determination of boiling range distribution一Gas chromatography method.
10.2 Sample analysis
Using the analysis sequence protocol (see 9.1), inject a sample aliquot into the gas chromatograph. At the
time of injection, start the chromatograph time cycle and the integrator/computer data acquisition.
11 Calculation
11.1 Correct the sample area slices for the non-sample detector response by subtracting each blank analysis area slice from each sample area slice at the equivalent slice time. Sum the corrected area slices to obtain the cumulative corrected areas for each time interval during the run.
11.2 At the point on the chromatogram where the baseline at the end of the run first becomes steady, record the total cumulative area counts. Move back along the chromatogram until the cumulative area equals 99,5 % of the total area. Mark this point as the final boiling point.
NOTE Locating the final boiling point can be the most difficult step in this method. Some samples have extremely long tail-end portions due to gradually decreasing amounts of heavy material. This fact, coupled with the natural tendency of the chromatographic baseline to rise at the end of the run due to septum and/or column bleed and/or elution of traces of heavy components from previous samples, can preclude the possibility of the chromatogram returning precisely to the original baseline established prior to the initial boiling point of the sample. Thus, the most satisfactory procedure is to inspect the chromatogram and the area counts at each interval near the end of the run to determine the point at which the rate of change of the chromatographic signal has reached a constant low value of no greater than 0,000 01 % of the total area counts per second.
11.3 Observe the area counts at the start of the run until the point is reached where the cumulative area count is equal to 0,5 % of the total area. Mark this point as the initial boiling point of the sample. If carbon disulfide is used as the solvent, its response shall be ignored in the calculations.
11.4 Divide the cumulative area at each interval between the initial and final boiling points by the total area and multiply by 100 to give the percentage of the sample recovered at each time interval.
11.5 Tabulate the cumulative percentage recovered at each interval and the retention time at the end of the interval. Using linear interpolation where necessary, determine the retention time associated with each percentage between 1 % and 99 %.
11.6 For each percentage and its associated retention time, determine the corresponding boiling point temperature from the calibration table (see 9.3.2). Use linear interpolation between data points.
12 Expression of results
12.1 Report the temperature to the nearest 0,5 °C at 1 % intervals between 1 % and 99 % and at the IBP and the FBP.
12.2 If a plot of the boiling-point distribution curve is required, use graph paper with uniform subdivisions and
plot each boiling temperature against its corresponding percentage recovered. Plot the initial boiling point at 0 % and the final boiling point at 100 % recovered. Draw a smooth curve connecting the points.
13 Precision
13.1 General
The precision, as determined by statistical examination in accordance with ISO 4259 of interlaboratory test results, is given in 13.2 and 13.3.
13.2
Repeatability
The difference between two test results, obtained by the same operator with the same apparatus under constant operating conditions on identical test material would, in the long run, in the normal and correct operation of the test method, exceed the values given in Table 7 in only one case in 20.

Main Focus Download

LEAVE A REPLY

Anonymous netizen Fill in information